ASM Failure Analysis Case Histories: Medical and Biomedical Devices
Fatigue Crack on a Type 316 Stainless Steel Bone Plate and Corresponding Broken Screw
-
Published:2019
Abstract
A narrow bone plate made of type 316 stainless steel and used to stabilize an open midshaft femur fracture failed. A crack at a plate hole next to the fracture site had been revealed by a radiograph taken 13 weeks after the operation. The plate was revealed to be slightly bent in the horizontal plane, and the fracture gap was considerably open. The screws and plates supplied by different manufacturers were revealed to be different with respect to microcleanliness (primary inclusion content) of the materials and only one of them was found to be according to specifications. The local...
Sign in
ASM members
Member Sign InFatigue Crack on a Type 316 Stainless Steel Bone Plate and Corresponding Broken Screw, ASM Failure Analysis Case Histories: Medical and Biomedical Devices, ASM International, 2019, https://doi.org/10.31399/asm.fach.med.c0048410