Skip Nav Destination
Close Modal
By
E. Proverbio, L.M. Bonaccorsi
By
Charlie R. Brooks, Ashok Choudhury
By
B. Aksakal, O.S. Yildirim, H. Gul
By
C.R.F. Azevedo, E. Hippert, Jr.
By
Cassio Barbosa, Ibrahim de Cerqueira Abud, Tatiana Silva Barros, Sheyla Santana de Carvalho, Ieda Maria Vieira Caminha
By
Harold J. Snyder, Casey B. Snyder
Search Results for
implant
Update search
Filter
- Title
- Authors
- Author Affiliations
- Full Text
- Abstract
- Keywords
- DOI
- ISBN
- EISBN
- Issue
- ISSN
- EISSN
- Volume
- References
Filter
- Title
- Authors
- Author Affiliations
- Full Text
- Abstract
- Keywords
- DOI
- ISBN
- EISBN
- Issue
- ISSN
- EISSN
- Volume
- References
Filter
- Title
- Authors
- Author Affiliations
- Full Text
- Abstract
- Keywords
- DOI
- ISBN
- EISBN
- Issue
- ISSN
- EISSN
- Volume
- References
Filter
- Title
- Authors
- Author Affiliations
- Full Text
- Abstract
- Keywords
- DOI
- ISBN
- EISBN
- Issue
- ISSN
- EISSN
- Volume
- References
Filter
- Title
- Authors
- Author Affiliations
- Full Text
- Abstract
- Keywords
- DOI
- ISBN
- EISBN
- Issue
- ISSN
- EISSN
- Volume
- References
Filter
- Title
- Authors
- Author Affiliations
- Full Text
- Abstract
- Keywords
- DOI
- ISBN
- EISBN
- Issue
- ISSN
- EISSN
- Volume
- References
NARROW
Format
Topics
Book Series
Date
Availability
1-20 of 92
Search Results for implant
Follow your search
Access your saved searches in your account
Would you like to receive an alert when new items match your search?
1
Sort by
Book Chapter
Microstructural Analysis of Failure of a Stainless Steel Bone Plate Implant
Available to PurchaseSeries: ASM Failure Analysis Case Histories
Publisher: ASM International
Published: 01 June 2019
DOI: 10.31399/asm.fach.med.c9001579
EISBN: 978-1-62708-226-6
... Abstract Stainless steel is frequently used for bone fracture fixation in spite of its sensitivity to pitting and cracking in chloride containing environments (such as organic fluids) and its susceptibility to fatigue and corrosion fatigue. A 316L stainless steel plate implant used for fixation...
Abstract
Stainless steel is frequently used for bone fracture fixation in spite of its sensitivity to pitting and cracking in chloride containing environments (such as organic fluids) and its susceptibility to fatigue and corrosion fatigue. A 316L stainless steel plate implant used for fixation of a femoral fracture failed after only 16 days of service and before bone callus formation had occurred. The steel used for the implant met the requirements of ASTM Standard F138 but did contain a silica-alumina inclusion that served as the initiation point for a fatigue/corrosion fatigue fracture. The fracture originated as a consequence of stress intensification at the edge of a screw hole located just above the bone fracture; several fatigue cracks were also observed on the opposite side of the screw hole edge. The crack propagated in a brittle-like fashion after a limited number of cycles under unilateral bending. The bending loads were presumably a consequence of leg oscillation during assisted perambulation.
Book Chapter
Metallurgical Failure Analysis of a Broken Compression Hip Screw Surgical Implant
Available to PurchaseSeries: ASM Failure Analysis Case Histories
Publisher: ASM International
Published: 01 June 2019
DOI: 10.31399/asm.fach.med.c9001664
EISBN: 978-1-62708-226-6
... Plastic deformation Surgical implants 316L UNS S31603 Fatigue fracture Corrosion fatigue In fractures of and around the head of the femur, a compression hip screw surgical implant is sometimes employed to provide bone alignment and fixation during healing. Figure 1 shows schematically...
Abstract
A compression hip screw is a device designed to hold fractures in the area of the femur in alignment and under compression. A side plate, which is an integral part of the device, is attached by screws to the femur, and it holds the compression screw in position. The device analyzed had broken across the eighth hole (of nine holes) from the end of the plate. The detailed metallurgical failure analysis of the device, including metallography and fractography, is reported here. It was found that the device had adequate metallurgical integrity for the application for which it was intended. It is believed that failure was caused by the lack of a screw in the ninth hole. Evidence is also presented which indicates that the device was bent prior to insertion, and the local plastic deformation may have caused structural changes leading to premature crack initiation.
Book Chapter
Metallurgical Failure Analysis of Various Implant Materials Used in Orthopedic Applications
Available to PurchaseSeries: ASM Failure Analysis Case Histories
Volume: 3
Publisher: ASM International
Published: 01 December 2019
DOI: 10.31399/asm.fach.v03.c9001803
EISBN: 978-1-62708-241-9
.... orthopedic implant devices fracture corrosion inclusions and stress gaps medical materials cracking fretting pitting fractography fracture toughness Ti6Al4V (titanium-aluminum-vanadium alloy) UNS R56406 316L stainless steel (austenitic wrought stainless steel) UNS S31603 Introduction...
Abstract
Metallurgical SEM analysis provides many insights into the failure of biomedical materials and devices. The results of several such investigations are reported here, including findings and conclusions from the examination a total hip prosthesis, stainless steel and titanium compression plates, and hollow spinal rods. Some of the failure mechanisms that were identified include corrosive attack, corrosion plus erosion-corrosion, inclusions and stress gaps, production impurities, design flaws, and manufacturing defects. Failure prevention and mitigation strategies are also discussed.
Image
Published: 01 January 2002
Fig. 39 Stress-corrosion cracking in a 316 stainless steel orthopedic implant
More
Image
Fatigue curves of type 316LR stainless steel implant material tested in ben...
Available to PurchasePublished: 01 January 2002
Fig. 21 Fatigue curves of type 316LR stainless steel implant material tested in bending mode. (a) S-N curves for stainless steel in cold-worked and soft condition that was tested in air and aerated lactated Ringer's solution. (b) Fatigue curve for number of cycles to failure as shown in Fig
More
Image
Designs of an asymmetric and a symmetric fixation implant device and the ap...
Available to Purchase
in Fatigue Failures of Austenitic Stainless Steel Orthopedic Fixation Devices
> ASM Failure Analysis Case Histories: Medical and Biomedical Devices
Published: 01 June 2019
Fig. 3 Designs of an asymmetric and a symmetric fixation implant device and the applied finite-element mesh with the von Mises stress contours for a unit stress applied to the cross section of the two plates
More
Image
Published: 15 January 2021
Fig. 39 Stress-corrosion cracking in a 316 stainless steel orthopedic implant
More
Image
in Failure of a Stainless Steel Hip Fracture Fixation Device
> Handbook of Case Histories in Failure Analysis
Published: 01 December 1993
Fig. 1 As-received Jewett nail implant.
More
Image
Underside of porous-coated implant that failed by fatigue. Two fatigue crac...
Available to Purchase
in Fatigue Fracture of Titanium Alloy Knee Prostheses
> Handbook of Case Histories in Failure Analysis
Published: 01 December 1993
Fig. 3 Underside of porous-coated implant that failed by fatigue. Two fatigue cracks are present: the failure and a secondary crack parallel to the fracture surface. Porous beads fractured in the region adjacent to each fracture and migrated into the joint cavity.
More
Image
in Fatigue Fracture of Titanium Alloy Knee Prostheses
> Handbook of Case Histories in Failure Analysis
Published: 01 December 1993
Fig. 4 Top surface of implant shown in Fig. 3 . Note the severe wear pattern and the presence of beads in the polyethelene surface.
More
Image
in Failure Analysis of Medical Devices
> Analysis and Prevention of Component and Equipment Failures
Published: 30 August 2021
Fig. 10 Schematic showing the “race to healing” for an implant
More
Image
in Failure Analysis of Medical Devices
> Analysis and Prevention of Component and Equipment Failures
Published: 30 August 2021
Fig. 17 Fracture surface of a fractured Ti-6Al-4V hip implant
More
Image
Optical microscopy image of fracture surfaces of a knee implant hinge post,...
Available to Purchase
in Failure Analysis of Medical Devices
> Analysis and Prevention of Component and Equipment Failures
Published: 30 August 2021
Fig. 20 Optical microscopy image of fracture surfaces of a knee implant hinge post, which fractured in vivo
More
Image
Porous, coated knee implant, as received. Insufficient bonding occurred at ...
Available to Purchase
in Failure of the Bond Between a Cobalt Alloy Prosthetic Casting and a Sintered Porous Coating
> Handbook of Case Histories in Failure Analysis
Published: 01 December 1992
Fig. 1 Porous, coated knee implant, as received. Insufficient bonding occurred at the tip of the patella flange.
More
Image
in Metallurgical Failure Analysis of Various Implant Materials Used in Orthopedic Applications
> Handbook of Case Histories in Failure Analysis
Published: 01 December 2019
Fig. 1 Several implant failures that occurred in the authors’ patients
More
Series: ASM Handbook Archive
Volume: 11
Publisher: ASM International
Published: 01 January 2002
DOI: 10.31399/asm.hb.v11.a0001819
EISBN: 978-1-62708-180-1
... Abstract This article commences with a description of the prosthetic devices and implants used for internal fixation. It describes the complications related to implants and provides a list of major standards for orthopedic implant materials. The article illustrates the body environment and its...
Abstract
This article commences with a description of the prosthetic devices and implants used for internal fixation. It describes the complications related to implants and provides a list of major standards for orthopedic implant materials. The article illustrates the body environment and its interactions with implants. The considerations for designing internal fixation devices are also described. The article analyzes failed internal fixation devices by explaining the failures of implants and prosthetic devices due to implant deficiencies, mechanical or biomechanical conditions, and degradation. Finally, the article discusses the fatigue properties of implant materials and the fractures of total hip joint prostheses.
Book Chapter
Retrieval and Analysis of Surgical Implants in Brazil: The Need for Proper Regulation
Available to PurchaseSeries: ASM Failure Analysis Case Histories
Publisher: ASM International
Published: 01 June 2019
DOI: 10.31399/asm.fach.med.c9001690
EISBN: 978-1-62708-226-6
... Abstract This paper summarizes several cases of metallurgical failure analysis of surgical implants conducted at the Laboratory of Failure Analysis of IPT, in Brazil. Investigation revealed that most of the samples were not in accordance with ISO standards and presented evidence of corrosion...
Abstract
This paper summarizes several cases of metallurgical failure analysis of surgical implants conducted at the Laboratory of Failure Analysis of IPT, in Brazil. Investigation revealed that most of the samples were not in accordance with ISO standards and presented evidence of corrosion assisted fracture. Additionally, some components were found to contain fabrication/processing defects that contributed to premature failure. The implant of nonbiocompatible materials results in immeasurable damage to patients as well as losses for the public investment. It is proposed that local sanitary regulation agencies create mechanisms to avoid commercialization of surgical implants that are not in accordance with standards and adopt the practice of retrieval analysis of failed implants. This would protect the public health by identifying and preventing the main causes of failure in surgical implants.
Book Chapter
Microscopic Analysis of Fractured Screws Used as Implants in Bone Fixation
Available to PurchaseSeries: ASM Failure Analysis Case Histories
Volume: 3
Publisher: ASM International
Published: 01 December 2019
DOI: 10.31399/asm.fach.v03.c9001804
EISBN: 978-1-62708-241-9
... Abstract A stainless steel screw securing an orthopedic implant fractured and was analyzed to determine the cause. Investigators used optical and scanning electron microscopy to examine the fracture surfaces and the microstructure of the austenitic stainless steel from which the screw was made...
Abstract
A stainless steel screw securing an orthopedic implant fractured and was analyzed to determine the cause. Investigators used optical and scanning electron microscopy to examine the fracture surfaces and the microstructure of the austenitic stainless steel from which the screw was made. The results of the study indicated that the screw failed due to fatigue fracture stemming from surface cracks generated by stress concentration likely caused by grooves left by improper machining.
Book Chapter
Fatigue Fracture of 316L Stainless Steel Screws Employed for Surgical Implanting
Available to PurchaseSeries: ASM Failure Analysis Case Histories
Volume: 1
Publisher: ASM International
Published: 01 December 1992
DOI: 10.31399/asm.fach.v01.c9001097
EISBN: 978-1-62708-214-3
... Abstract Two type 316L stainless steel orthopedic screws broke approximately 6 weeks after surgical implant. The screws had been used to fasten a seven-hole narrow dynamic compression plate to a patient's spine. The broken screws and screws of the same vintage and source were examined using...
Abstract
Two type 316L stainless steel orthopedic screws broke approximately 6 weeks after surgical implant. The screws had been used to fasten a seven-hole narrow dynamic compression plate to a patient's spine. The broken screws and screws of the same vintage and source were examined using macrofractography, SEM fractography, and hardness testing. Fractography established that fracture was by fatigue and that the fatigue cracking originated at corrosion pits. Hardness while below specification, still indicated that the screws were in the cold-worked condition and notch sensitive during fatigue loading. Use of a steel with a higher molybdenum content (317L) in the annealed condition was recommended.
Image
Effect of ion implantation on the coefficient of friction in fretting of IM...
Available to PurchasePublished: 01 January 2002
Fig. 28 Effect of ion implantation on the coefficient of friction in fretting of IMI 550 titanium alloy at 500 °C (930 °F)