ASM Failure Analysis Case Histories: Medical and Biomedical Devices
Titanium Screw Head With Fretting Structure at Contact Area With Plate Hole
-
Published:2019
Abstract
Wear on a titanium screw head with a lip of material that that was transported by fretting at a plate-hole edge was studied. A flat fretting zone was visible on the screw surface over the material lip. A cellular wear structure containing wear debris was found. No morphological signs of corrosion were observed in connection with fretting structures.
Titanium Screw Head With Fretting Structure at Contact Area With Plate Hole, ASM Failure Analysis Case Histories: Medical and Biomedical Devices, ASM International, 2019, https://doi.org/10.31399/asm.fach.med.c0048421