Skip to Main Content

Abstract

During the internal fixation, the type 316LR stainless steel cortical bone screw failed. Extensive spiral deformation was revealed by the fracture surface. Dimple structure characteristic of a ductile failure mode was observed with dimples oriented uniformly in the deformation direction. A zone of heavily deformed grains at the fracture edge was revealed by longitudinal metallographic examination. The shearing fractures of a commercially pure titanium screw and a cast cobalt-chromium-molybdenum alloy were discussed for purpose of comparison.

You do not currently have access to this chapter.
Don't already have an account? Register

2019. "Shearing Fracture of a Type 316LR Stainless Steel Screw", ASM Failure Analysis Case Histories: Medical and Biomedical Devices